OMA Metal Ion Analyzer

OMA Metal Ion Analyzer


METAL ION ANALYZER

OMA continuously analyzes liquid samples and outputs real-time concentrations of metal ions, including copper, iron, nickel, chromium, cobalt, and others.

Product Features


Performance

Rapid, consecutive readings T90 response time is less than 10 seconds. No false alarms Not affected by the interference of potentially other chemical substances. Extremely wide dynamic measurement range It can maintain high accuracy even when the sample concentration varies significantly. Fully automated operation under unmanned supervision Automatic purging and zeroing function; optional remote automated control

Reliability

Fully automated operation under unmanned supervision Automatic purging and zeroing function; optional remote automated control Inherent Security Design The sample never comes into contact with or enters the interior of the analyzer housing. Customize various alarm and communication outputs. 4-20mA analog output, MODBUS; control via HART and other protocols

Flexibility

Can directly analyze hot/humid samples. The standard flow cell can withstand samples at 150° and 3000 psi. It can simultaneously measure five types of metal ions. Use powerful multi-component online monitoring software.

Usage cost

No consumable materials or reagents. In addition to purging with standard gas, Low maintenance requirements Simply require wiping the lenses once a month. Long-life xenon light source Average service life of 5 years Instrument with an all-solid-state structure Stands out among competitors due to the absence of moving parts.

Compatibility

"Clean Flange"-type sample flow cell with a specially designed structure Designed for environmental sanitation applications Explosion-proof certificate Class I Div 1 & 2; ATEX; GOST... View certifications No-analysis cabin Can be installed directly outdoors.

Technical concept


diode-array-400w.webp

Monitoring Principle

The OMA online analyzer uses the principles of ultraviolet and visible spectroscopy to determine the concentrations of chemical components in a sample. The OMA system analyzes the absorbance of the sample across wavelengths from 200 to 800 nanometers, and the data is captured by a 1024-element photodiode array detector and transmitted to a processor for data analysis, yielding high-resolution absorption spectra.

oma-optical-assembly-400w.webp

Optical structure

The OMA process online analyzer uses a long-life xenon lamp as its light source to transmit signals that come into contact with the sample via a flow cell. This signal is emitted by the light source and transmitted through an optical fiber cable into the flow cell, where it interacts with the sample gas. Samples composed of various chemical substances exhibit unique absorption characteristics for light. The resulting light signal then exits the flow cell, travels back to the detector via another optical fiber cable.

metal-ions-spectra.webp

Absorbance Curve: Metal Ions

To detect the concentration of metal ions, OMA identifies the unique absorbance spectrum of pure metal ions and uses mathematical methods to separate and isolate the absorbance spectrum of the metal ions from that of the sample being analyzed. Based on Beer-Lambert’s law, OMA correlates the unique absorbance spectrum of the metal ions with their actual concentrations.

oma-nh3-software-400w.webp

User interface

AAI’s proprietary ECLIPSE software processes the acquired spectral data into actual concentration readings. On the touchscreen controller, the analyzer operator can easily switch between different interfaces—such as trend graphs, spectra plots, concentration readings, and more—and can independently and at any time set and adjust alarms, outputs, purge functions, zeroing, and other parameters.

analyzer-safety-banner-400w.webp

Safety performance

Most analyzers directly introduce the sample into the analyzer’s housing for measurement. However, if the sample is toxic, highly corrosive, or explosive, this could be extremely dangerous. The OMA design is unique: instead of introducing the sample into our analyzer, we guide the optical signal into the sample itself. The sample flows only through a flow cell and never comes into contact with the analyzer. The flow cell is connected to the analyzer via an optical fiber cable.