Product
Contact
Beijing Qinglian Measurement and Control Technology Co., Ltd
Tel:18500242456
Email:sales@qingliancekong.com
Address: Inside Beijing Taixing Art Institute, Shuipo Village, Gaoliying Town, Shunyi District, Beijing
TDL-406
Rack-mounted laser gas analyzer
The TDL-406 is a tunable diode laser analyzer capable of real-time, online analysis of a variety of different gases. The analyzer can be mounted in a 19-inch rack and is well-suited for integration into analytical cabinets as well as laboratory environments.
Product Features
Measurement range
- Ammonia NH 3 – Low range: 0-20 ppm; High range: 0-100 ppm
- Hydrogen chloride HCl – Low range: 0–50 ppm; High range: 0–100 ppm
- Water H 2 O – 0-30%
- Methane CH 4 – 0-100 ppm
- Carbon dioxide CO 2 – 0-1000 ppm
Product Features
- Measurement method in which optical components do not come into contact with the sample gas
- Precision laser diode optoelectronic components
- High selectivity
- Fast response time
- Can be installed independently
- Continuous sensor status monitoring
- Low maintenance requirements
- Low holding costs
- Digital Communication and Analog Communication Interfaces
- Can measure gases up to 190 degrees Celsius.
Technical drawing
Technical concept
Technical Parameters
The following technical specifications are based on the premise that all components and the sample system are supplied by AAI, and that the system installation has been approved by AAI. For technical specifications under other conditions, please consult AAI’s sales team directly.
General
| Measurement principle | Tunable Diode Laser Spectroscopy (TDLS) |
|---|---|
| Sample medium | Gaseous |
| Instrument calibration | Calibration before shipment |
| Measurement verification | Zero-point calibration and standard gas calibration |
| Controller | U.S. AAI Standard Controller: Industrial machine controller with an 8-inch LCD touchscreen. Data table |
| Software | ECLIPSE TM Online TDL |
| Data storage | Solid-state drive. Data table |
| Chassis | 19-inch 5U rack-mountable |
| Operating environment | |
| Analyzer environmental conditions | Indoor or in a cabinet |
| Ambient temperature | 15–50 degrees Celsius |
| Dimensional parameters | |
| Instrument size | 8.64 inches high × 19.00 inches wide × 14.05 inches deep |
| Weight | 30 pounds |
| Process gas connection size | 1/4 inch |
| Public facility requirements | |
| Power requirements | 100-240 VAC |
| Power consumption | 370 watts |
| Output and Communication | |
| Output | 4-20mA analog output along the way |
| Continuous fault alarm (dry contact) | |
| Ethernet-based remote control connection for the entire route | |
| Modbus serial link options available along the way | |